The GS820 Multi-Channel Source Meter comes in 18-volt and 50-volt models for testing LED lighting, measuring drain current, and more. It has two isolated analog channels, each with a constant voltage source, constant current source, voltmeter, and ammeter. The Yokogawa source measure unit, GS820 offers two sweep functions, PC connectivity via USB, and expansion up to 10 channels.
Applications:
- DC voltage/current reference
- I/V curve trace
- Semiconductor parametric test
- Pulse voltage/current source
- Electronic load
- Resistance measurement
- Programmable arbitrary
- Production test waveform generation
Source and Measurement Functions:
- Voltage source and current measurement (VS&IM)
- Current source and voltage measurement (IS&VM)
- Voltage source (VS)
- Current source (IS)
- Voltmeter (VM)
- Ammeter (IM)
- Resistance meter (IS&VM)
These functions can be selected for each channel to form an arbitrary combination of functions.
Four-quadrant operation consisting of source operation (current source) and sink operation (current sink) is available with ranges up to 7 V and 3.2 A or 18 V and 1.2 A. The output and measurement resolutions are 5.5 digits. Other features include:
- Isolated 2-channel source and measurement function
- Source and measurement ranges:
18V model: 7 V and 3.2 A or 18 V and 1.2 A
50V model: 20 V and 1.2 A or 50 V and 0.6 A
- 1 pA resolution at extremely small current range 200 nA
- Generate arbitrary waveforms consisting of up to 100,000 points at 100 μs intervals
- Channel expansion through master-slave synchronization link
- Fast test speeds
- 16-bit digital I/O (model 765602/12)
Source and Measurement Range
Four-quadrant operation consisting of source operation (current source) and sink operation (current sink) is available. The output and measurement resolutions are 5.5 digits. Two models are available for your application.
18 V model (765601/02)
- Voltage ranges:
- 200 mV, 2 V, 7 V, and 18 V
- Maximum output current:
- ±3.2 A (at an output voltage of ±7 V or less) ±1.2 A (at an output voltage of ±18 V or less)
- Current ranges:
- 200 nA, 2 µA, 20 µA, 200 µA, 2 mA, 20 mA, 200 mA, 1 A, and 3 A
- Maximum output voltage:
- ±18 V (at an output current of ±1.2 A or less) ±7 V (at an output current of ±3.2 A or less)
50 V model (765611/12)
- Voltage ranges:
- 200 mV, 2 V, 20 V, and 50 V
- Maximum output current:
- ±1.2 A (at an output voltage of ±20 V or less) ±0.6 A (at an output voltage of ±50 V or less)
- Current ranges:
- 200 nA, 2 µA, 20 µA, 200 µA, 2 mA, 20 mA, 200 mA, 0.5 A, and 1 A
- Maximum output voltage:
- ±50 V (at an output current of ±0.6 A or less) ±20 V (at an output current of ±1.2 A or less)